Cart0Item(s)

You have no items in your shopping cart.

Product was successfully added to your shopping cart.

Advanced 12 inch portable Patient Monitor MSLMP03

Quick Overview

Brand:MedsinglongService:Medsinglong after-sales support.Contact us: Cindy@medmsl.com

Availability: In stock

$9,999.00

Details

Advanced 12 inch portable Patient Monitor MSLMP03 Multi-parameter(ECG,SPO2,RESP, Pulse rate, TEMP,NIBP)

  • Portable Patient Monitor MSLMP03 Multi-parameter is a robustpatient monitor,small enough to move around,yet flexible enough to adapt to your particular needs.
  • With a variety of optional extras,including arrhythmia and ST analysis,EtCO2 measurement and double IBP,MSLMP03 multi-parameter patient monitor is configurable to suit your requirements.
  • Purchase only the advan ced features that you need and nothing more.With NIBP recall and extensive data storage capability,Portable patient Monitor MSLMP03 Multi-parameteris particularly well-suited for use in peri-anesthesia and critical care environments.
  • Rich in functions,conveniently adaptable and pleasingly affordable,Portable patient Monitor MSLMP03 Multi-parametermay be just the monitor for your department.

portable Patient Monitor Features:

  • *12" color TFT display,up to 12 waveforms
  • *One level menu,easy to use
  • *Parameters including ECG/RESP,SpO2 ,NIBP,TEMP,Pulse Rate and IBP(optional)
  • *Maximum 720 hour graphic and tabular trends of all parameters
  • *Built*in recorder and battery
  • *Suitable for adult,pediatric and neonatal patient
  • *Arrhythmia analysis,ST calculation,Pace analysis
  • *Anti*ESU,anti-defibrillator
  • *Standard network interface,can connect to Central monitor system
portable Patient Monitor Main Functions:
  • 1.12" widescreen TFT,Resolution ratio: 800×600,up to 12 waveforms ;
  • 2.Display ECG,SpO2,pulse rate,NIBP,respiration,temperature
  • 3.Plumbic acid battery.
portable Patient Monitor Optional:
  • 1.Recorder(Built*in)
  • 2.CO2
  • 3.Anesthesia depth

MSLMP03 multi-parameter patient monitor specifications

  • 1.Classification
    • SFDA classification:    Class II
    • Anti-electroshock degree:    Class I equipment with internal power supply
    • Anti-electroshock degree    
    • TEMP/SpO2/NIBP:     BF
    • ECG/RESP:    CF
    • Explosion proof level:    Ordinary equipment,without explosion proofm
    • Harmful liquid proof degree:    Ordinary equipment,without liquid proof
    • Working system:    Continuous running equipment
  • 2.applicable standards
    • Medical electrical equipment GB 9706.1-2007,Part I:General requirements for safety
    • GB 9706.25-2005 Medical electrical equipment,Part 2-27:ECG monitoring equipment safety requirements
    • YY 1079-2008 ECG
    • YY0667-2008 Medical electrical equipment Part 2:Automatic cycling non-invasive blood pressure monitoring equipment,safety and essential performance.
    • YY0668-2008 Medical electrical equipment Part 2:multi-parameter patient monitoring equipment requirements for the safety.

  • 3.monitor size and weight
    • size:300 ×275 ×150 mm  
    • Weight:4 kg
  • 4.Power Supply:100~240 VAC,50/60 Hz,Pmax=90VA Fuse T1.5A
  • 5.Battery
    • 2.6 Ah 12V rechargeable battery
    • Operating time after full charge is more than 1 hours
    • Operating time after the first alarm of low battery will be about 5 minutes
    • Maximum charging time is less than 6 hours.
  • 6 Signal Interface:Network interface  standard RJ45 Socket
  • 7 Storage
    • Trend:720 hours
    • NIBP review:1000 NIBP events
    • Wave review:2 hours
    • Alarm review:200 alarm events
    • All storage data are non volatile.
  • 8 Environment
  • 8.1 Temperature
    • Working:0~40 °C
    • Storage:-20~50°C
  • 8.2 Humidity
    • Working:15%-90 % (no coagulation)
    • Storage:15%-90 % (no coagulation)
  • 8.3 Atmospheric pressure
    • Working 86.0 kPa~106.0kPa;
    • Storage 86.0 kPa~106.0kPa
  • 9 ECG Specification
  • 9.1 Heart rate calculation method
    • Since the first occurrence of heart rate to heart rate occurred in the second interval T1;second heart rate,heart rate occurred to a third place time interval T2;continuous record 10 times,and then calculated the average interval of 10,must the heart rate per minute.3s~10s refresh every heart.
  • 9.2 Heart rate meter accuracy and arrhythmia response
    • Monitor the 20s stable time after figure ECG heart rate a wave group):heart rate display 40bpm±5bpm;b):heart rate display 30bpm±5bpm;c):heart rate display 120bpm±5bpm;d):heart rate display 45bpm±5bpm
    • A) Couple rhythm—two waves of duration is 1500ms;if the calculation all the QRS complex,heart rate to 80bpm,if only large R wave or S-wave,the heart rate to 40bpm.
    • B) Slow change couple rhythm—if the calculation all the QRS complex,heart rate to calculate 60bpm,if only large waves,heart rate to 30bpm.
    • C) Fast couple rhythm—if the calculation all the QRS complex,heart rate as 120bpm.
    • D) bi-directional contraction—if you calculate the QRS complex,heart rate to calculate 90bpm,if only large waves,heart rate to 45bpm.
  • 9.3 Lead mode
    • 5 Leads:RA、LA、LL、RL、V;lead mode:I,II,III,AVR,AVL,AVF,V
  • 9.4 Gain
    • Gain control:manually four stalls
    • 1/4 stalls(×0.25) 2.5mm/mV;1/2 stalls (×0.50) 5.0mm/mV; 1 stall (×1) 10mm/mV;2 stalls (×2) mV 20mm/mv;
  • 9.5 Heart rate
    • Measure range:
    • Adult:15~300 bpm
    • Neonatal/Pediatric:15~350 bpm
    • accuracy:±1%
    • resolution:1 bpm
  • 9.6 Sensitivity
    • >200 μV P-P
  • 9.7 Differential Input Impedance
    • >5 M ohm
  • 9.8 Bandwidth
    • Surgery:1~20 Hz
    • Monitor:0.5~40 Hz
    • Diagnostic;0.05~130 Hz
  • 9.9 CMRR
    • Diagnostic Mode >90 dB
    • Monitor Mode:>110 dB
    • Surgery Mode:>110 dB
  • 9.10 Electrode offset potential:±300mV
  • 9.11 Input dynamic range
    • In any lead,the applied DC bias voltage of±300mV to 320mV/s rate of change of±5mV of the differential mode voltage,and display devices should have the capacity to respond;output signal amplitude error should not exceed±10% ;
  • 9.12 pacing pulse suppression
    • Monitor switch is open in the pacemaker,can inhibit the range:±2 mV~±700mV,width:0.1ms~2ms,rise time:10us~100μs normal QRS wave single pulse without overshoot pacing pulse does not affect heart rate calculation.
    • Monitor switch is open in the pacemaker,can inhibit the range:±2 mV~±5mV,width:0.5ms~2ms,rise time:10us~100μs single pulse overshoot normal QRS waves of pacing pulses,without affecting heart rate calculation.
    • Overshoot (a0) range should 0.025ap to 0.25ap range,independent of the choice of time constant,but not more than 2mV;
    • Monitor switch is open in the pacemaker,pacing pulse detector inhibition of rapid ECG signal to 1V/s RTI minimum input slew rate.
  • 9.13 QRS wave amplitude and period range between
    • Amplitude (p-v RTI) range:0.5mV~5mV;
    • Width (adult monitor):70~120ms;
    • Width (neonatal/pediatric):40~120ms;
    • Set monitor mode for adults should not be less than or equal to 0.15mV amplitude QRS signal or of less than or equal to 10ms between the 1mV signal a response.For neonatal/pediatric mode,monitors allowed to respond to the above two signals to respond to either or both;
  • 9.14 frequency voltage tolerance
    • Superimposed on a QRS wave signals received,and to meet tolerances of±10% of heart rate showed the error limits of accuracy required under the premise of the maximum frequency sine wave peak-valley kurtosis should not be less than 100μV (pv)
  • 9.15 drift tolerance
    • When a 0.1Hz,amplitude 4mV (pv) of the triangular wave superimposed on a string of rate of 0.5mV,interval 100ms,repetition rate of 80bpm when the QRS wave,ECG heart rate monitor shows the error should be within 80bpm±8bpm;
  • 9.16 baseline control and stability
    • Reset:reset recovery time is not greater than the 3s;
    • Baseline Stability:After boot,10s baseline drift in the output rate should not exceed 10μV/s RTI;
    • After boot,1h,total drift should not exceed 500μV/s RTI;
    • Working temperature should not exceed 50μV/℃
  • 9.17 system noise
    • Should be not more than 30μV (p-v RTI)
  • 9.18 Multi-channel crosstalk
    • Should not be imposed by the channel signal is not imposed without the signal channel output generated more than spin plus signal (multiplied by the gain) 5%;
  • 9.19 Electro surgery interference suppression
    • Heart rate should not exceed the interference is not activated when the heart rate±10%,interfere with the duration should not exceed 10S;
  • 9.20 pacemaker pulse display capabilities
    • Show an increase is not less than 0.2mV RTI;in 10s,the baseline drift is not greater than 10mm;
  • 9.21 heart rate response time
    • Step increase in heart rate from 80bpm to 120bpm maximum response time of less than 10 s;
    • Heart rate from 80bpm to 40bpm step to reduce the maximum response time of less than 10 s;
  • 9.22 to the alarm time
    • High heart rate alarm start time is less than 10 s by YY 1079-2008;
    • Low heart rate alarm start time is less than 10 s according to YY 1079-2008;
    • Cardiac stop alarm start time is less than 10 s by YY 1079-2008;
    • Tachycardia alarm start time is less than 10 s.
  • 9.23 calibration signal
    • 1mV (peak to peak),the accuracy  ≤±5%
  • 10 NIBP Specifications
  • 10.1 Measurements : Pulse wave oscillation
  • 10.2 Mode:  Manual/Automatic/Continuous
  • 10.3 Automatic measurement mode of measurement interval
    • 1 minute,2 minutes,3 minutes,4 minutes,5 minutes,10 minutes,15 minutes,30 minutes,60 minutes,90 minutes,2 hours,3 hours,4 hours,8 hours
  • 10.4 Continuous measurement mode of measurement time : 5 minutes
  • 10.5 pulse rate range: 40-240 bpm
  • 10.6 Measurement range and accuracy
    • Range
    • 40~280mmHg systolic blood pressure for adults
    • Diastolic blood pressure 10~220mmHg
    • The average pressure of 20~240mmHg
    • 40~220mmHg systolic blood pressure of children
    • Diastolic blood pressure 10~160mmHg
    • 20~170mmHg mean pressure
    • Newborn systolic blood pressure 40~135mmHg
    • Diastolic blood pressure 10~100mmHg
    • The average pressure of 20~110mmHg
    • Static pressure range 0~300mmHg
    • 3mmHg static pressure±Accuracy
    • 5mmHg;±Blood pressure accuracy:the maximum average error of
    • Maximum standard deviation 8mmHg
  • 10.7 overvoltage protection
    • Adult model 300 mmHg
    • Child model 240 mmHg
    • Neonatal model 150 mmHg
  • 11 SpO ± Specifications
  • 11.1 oxygen saturation
    • Measuring range 0 to 100%
    • Resolution 1%
    • 2%±Accuracy of 70~100%:
    • <69%:not defined
  • 11.2:pulse rate
    • Range 20~250bpm
    • Resolution 1bpm
    • 3bpm±Accuracy
  • 12:temperature specifications
    • Sensor type 10K series,2.25K Series
    • Number of channels 2 channels
    • C°Range 0~50
    • C°Resolution of 0.1
    • C°(sensor error not included) 0.1° Accuracy
  • 13 recorder (optional)
    • Record width of 48 mm
    • Walking paper speed 25/50 mm/s
    • 2-channel waveform tracings

 

Product Detail and Usage Pictures - Advanced 12 inch portable Patient Monitor MSLMP03

Advanced 12 inch portable Patient Monitor MSLMP03Advanced 12 inch portable Patient Monitor MSLMP03Advanced 12 inch portable Patient Monitor MSLMP03Advanced 12 inch portable Patient Monitor MSLMP03Advanced 12 inch portable Patient Monitor MSLMP03Advanced 12 inch portable Patient Monitor MSLMP03Advanced 12 inch portable Patient Monitor MSLMP03Advanced 12 inch portable Patient Monitor MSLMP03Advanced 12 inch portable Patient Monitor MSLMP03

Contact US for more videos about: Advanced 12 inch portable Patient Monitor MSLMP03

Contact Information

  • *
  • *
  • *
  • *
  • *
  • *

Product Tags

Other people marked this product with these tags:

Use spaces to separate tags. Use single quotes (') for phrases.

Connect Alwaysx

Connect us in Whatsapp or Facebook Messenger

Messenger

whatsapp